Horizontal Rectangular Autoclave

CONSTRUCTION: The chamber is constructed of heavy-duty stainless steel 316 & thickness 5 to 6 mm and jacket constructed of 5 mm SS 304 & Outer SS 304 2 mm to withstand the pressure of 2.2 Kg/cm2 Corresponding to the temperature of 134C and also to withstand vacuum of -500 to -700 mm Hg. In Chamber. one rail is provided in the chamber bottom & two carriage trolley for easy loading & unloading for the material.

BOILER: The autoclave Boiler constructed with a heavy gauge SS 316 electrically operated. The thickness of a 3mm stainless steel Sheet suitably located under the sterilizer. High-grade flanged type heating elements are fitted from the front side for Easy replacement/repair. Fitted with water gauge glass for water level, safety valve, water inlet, and drain valves. It is Also provided with a Low Water Level cut-off device. This device protects the heater in the generator from burning out Dry. The steam pressure in the boiler is indicated on an individual pressure gauge.

DOORS: The Sterilizer is supplied with single door or double doors made of SS 316 The doors are having automatic Hinge type Arrangement. The doors will have unique arrangement to open them manually in case of power failure. The door Gasket Will be of Silicon food grade. Door thickness is 16 to 20mm Ss316

SAFETY FEATURES: Special provision prevents the operator from opening the door in following events:-
a) When the chamber is under pressure door is not open.
b) For double door type autoclave only one door
can be open at same time.
c) Vacuum Breaker.
d) Heavy duty safety valves for over pressure.
e) Water level glass tube for water level monitoring.

FITTINGS & PIPING: All pipe line is complete SS 304 and thickness & size as per design of machine. VACUUM SYSTEM: This system is used for effective air removal from the Autoclave chamber. It consists of an efficient water-ring-type Vacuum pump as per machine design. AUTOMATIC PROCESS CONTROL UNIT: The control unit is based on advance HMI With PLC controller technology. This unit control Different sterilization cycles With the help of in-built different program combinations. The control unit is provided With a manual operation facility And gives the status of live operation details.

TEMPERATURE AND PRESSURE:
Sterilizing Temp.: 121°C.
to134°C. Sterilizing Pressure: 1.2 to 2.2 kg/cm2, (15 psi to 30 psi)

OPERATING CYCLE: 121°C Sterilization Cycle 134°C Sterilization Cycle VLT Test Cycle Bowie dick test